TY - JOUR AU - Tomer, Anil K AU - Verma, Vanita AU - Bhateja, Nitin AU - Tirupathi, Pudu AU - Kumari, Sushma PY - 2021/01/17 Y2 - 2024/03/29 TI - TO COMPARE AND EVALUATE THE CYCLIC FATIGUE RESISTANCE OF THREE NITI ROTARY FILE SYSTEM-AN INVITRO STUDY JF - International Journal of Medical Science And Diagnosis Research JA - ijmsdr VL - 5 IS - 1 SE - Articles DO - 10.32553/ijmsdr.v5i1.743 UR - https://www.ijmsdr.com/index.php/ijmsdr/article/view/743 SP - AB - <p><strong>Aim:</strong>&nbsp; To compare and evaluate the cyclic fatigue resistance of three NiTi rotary file system</p><p><strong>Materials and Methods:</strong> Three NiTi rotary system (Hyflex EDM, Protaper Gold and One curve) were used in study . All files were tested in simulated constructed apparatus with angle of curvature 60<sup>o </sup>and radius of curvature 5mm.The rotary files were used in endomotor X smart (Densply) .All the instruments were utilized until fracture occurred, and then the number of cycles to failure (NCF) was calculated. The data were analyzed statistically using Annova test. The statistical significance level was set at P &lt; 0.05.</p><p><strong>Result:</strong> HEDM instruments had the highest cyclic fatigue resistance among all the other instruments (P &lt; 0.05). The one curve instruments had a significantly higher fatigue resistance than the Protaper gold instruments (P &lt; 0.05).</p><p><strong>Conclusion: </strong>Within the limitations of this in vitro study, it was found that the cyclic fatigue resistance was higher for the HEDM instruments than for the, Protaper gold and one curve instruments.</p><p><strong>Keywords:</strong> Cyclic fatigue resistance, heat?treated NiTi instruments, HyFlex EDM, OneCurve</p> ER -